The instrument should be in standby mode (indicated by the radial button in the microwave bridge control window and the top status bar in the top left of the main window). Verify that the orange Tank Full light is onĤ 3. Turn on heat exchanger at switch near the back wall of the room. Turn on the water source at the wall on the left side of the room. Turning on the instrument Microwave Bridge Heat Exchanger Power Supply Console 2. This will dissolve some of the components of the cavity. If you break a sample in the cavity, do not use solvent to wash the inside of the cavity. After connecting the new cavity, raise the iris screw to approximately the midpoint before attempting to auto tune. When changing cavities, always completely lower the iris screw before disconnecting the old cavity. When changing accessories or cavities, turn off the magnet. Always change to tune mode, wait for the instrument to change, then switch to standby or operate. Do not switch the instrument directly from standby to operate (or operate to standby). Always change samples while in standby mode. 2 General Notes Never look into the cavity or waveguide while the microwave bridge is on.